![Fluke Biomedical ProSim 8 Series User Manual Download Page 80](http://html1.mh-extra.com/html/fluke-biomedical/prosim-8-series/prosim-8-series_user-manual_2310745080.webp)
ProSim 8
Users Manual
68
Table 14. Autosequences (cont.)
Autosequence
Sequence Steps
Run Time
NIBP Sequence
Waveform: Square; Rate: 2.0 Hz; Amplitude: 1.0 mV
00:05
Waveform: Sine; Rate: 0.05 Hz; Amplitude: 1.0 mV
00:05
Waveform: Sine; Rate: 0.5 Hz; Amplitude: 1.0 mV
00:05
Waveform: Sine; Rate: 1 Hz; Amplitude: 1.0 mV
00:05
Waveform; Sine; Rate: 10 Hz; Amplitude: 1.0 mV
00:05
Waveform: Sine; Rate: 50 Hz; Amplitude: 1.0 mV
00:05
Waveform: Sine; Rate: 60 Hz; Amplitude: 1.0 mV
00:05
Waveform: Sine; Rate: 100 Hz; Amplitude: 1.0 mV
00:05
Waveform: Sine; Rate: 150 Hz; Amplitude: 1.0 mV
00:05
Waveform: Triangle; Rate: 0.125 Hz; Amplitude: 1.0 mV
00:05
Waveform: Triangle; Rate: 2.0 Hz; Amplitude: 1.0 mV
00:05
Waveform: Pulse; Rate: 60 BPM; Amplitude: 1.0 mV
00:05
REPEAT
Total time 01:10
IBP Sequence
IBP 1: 120/80 mmHg (Art), 2: 120/0 mmHg (LV)
01:00
IBP 1: 25/0 mmHg (RV), 2: 120/80 mmHg (Art)
01:00
IBP 1: 25/10 mmHg (PA), 2: 10/2 mmHg (PAW)
01:00
REPEAT Total
Time
03:00
Temperature Sequence
Temperature 30.0
°
C 00:20
Temperature 34.0
°
C 00:20
Temperature 37.0
°
C 00:20
Temperature 42.0
°
C 00:20
STOP Total
Time
01:20
View the Steps of an Autosequence
To view an autosequence:
1. Push
.
2. Push
,
,
, or
to highlight
Autosquences
.
3. Push
.
4. Push
or
to highlight an autosequence in the list. See Table 14.
5. Push
.
Summary of Contents for ProSim 8 Series