Vital Signs Simulator
Detailed Specifications
89
Non-Invasive Blood Pressure
Pressure Units
....................................................... mmHg or kPa
Manometer (Pressure Meter)
Range ................................................................. 0 mmHg to 400 mmHg
Resolution ........................................................... 0.1 mmHg
Accuracy .............................................................
±
(0.5 % r0.5 mmHg)
Pressure Source
Target Pressure Range ...................................... 15 mmHg to 400 mmHg
Resolution ........................................................... 1 mmHg
NIBP Simulations
Pulse ................................................................... 2 mmHg max into 500 ml NIBP system
Volume of air moved ........................................... 1.25 ml max
Simulations ......................................................... Systolic/diastolic (MAP)
Adult ................................................................ 60/30 (40), 80/50 (60), 100/65 (77); 120/80 (93); 150/100 (117); and
200/150 (167) and 255/195 (215)
Neonatal ......................................................... 35/15 (22); 60/30 (40); 80/50 (60);100/65 (77);120/80 (93) and 150/100
(117)
Pressure variability ......................................... Systolic and diastolic pressures are variable by 1 mmHg
Repeatability ....................................................... Within
±
2 mmHg (at maximum pulse size independent of device under
test)
Synchronization
Normal Sinus heart rates ................................ 30 BPM to 240 BPM
Maximum rate at 1 ml ................................. 240 BPM achievable with pulses up to 1 ml
Maximum rate at 1.25 ml ............................ 180 BPM
Arrhythmias ..................................................... Premature atrial contraction (PAC), Premature ventricular contraction
(PVC), atrial fibrillation, and missed beat.
Leak Test
Target Pressure .................................................. 20 mmHg to 400 mmHg
Elapsed time ....................................................... 0:30 minutes:seconds to 5:00 minutes:seconds in 30 second steps
Range ................................................................. 0 mmHg/minute to 200 mmHg/minute
Internal Leak rate .................................................... <2 mmHg/min into 500 ml rigid volume
Pressure Relief Test Range
................................. 100 mmHg to 400 mmHg
Oximeter SpO2 Optical Emitter and Detector (optional)
% O
2
Range ................................................................. 30 % to 100 %
Resolution ........................................................... 1 %
Accuracy
With oximeter manufacturer’s R-curve
Saturation within UUT specific range ..........
±
(1 count + specified accuracy of the UUT)
Saturation outside UUT specific range ....... monotonic with unspecified accuracy
With Fluke Biomedical R-curves
91 % to 100 % ............................................
±
(3 specified accuracy of the UUT)
81 % to 90 % ..............................................
±
(5 specified accuracy of the UUT)
71 % to 80 % ..............................................
±
(7 specified accuracy of the UUT)
Below 71 % ................................................. monotonic with unspecified accuracy
Heart Rate
Range ................................................................. 30 BPM to 300 BPM in 1 BPM steps. Oximeter SpO2 Optical Emitter
and Detector is synchronized with ECG rate delayed by 150 ms.
Accuracy ............................................................. ±1 % of setting