![Fluke Biomedical ProSim 8 Series User Manual Download Page 47](http://html1.mh-extra.com/html/fluke-biomedical/prosim-8-series/prosim-8-series_user-manual_2310745047.webp)
Vital Signs Simulator
Invasive Blood Pressure Simulation and Tests
35
You can go back a step when you push the
Previous
softkey. The Product
simulates that step for a full 15 seconds before it does the subsequent step.
When you push the
Previous
softkey while paused, the Product goes back a
step, but stays paused and sets the time to 15 seconds.
When you push the
Stop
softkey, the procedure simulation is stopped and the
initial test screen shown in Figure 19 shows on the display.
Simulate a Cardiac Catheterization Procedure
The Product simulates blood pressure measurements on both sides of a heart
valve. The pressure difference, or gradient across the valve, is used to determine
heart valve condition.
To simulate a Cardiac Catheterization procedure:
1. Push the
Tests
softkey in the IBP screen.
2. Push
or
to highlight
Cardiac Catheterization
.
3. Push
. The valve selection shows on the display.
To change the valve value:
1. Push
.
2. Push
or
to highlight
Aortic
,
Pulmonary
, or
Mitral
.
3. Push
to go back to the IBP: Cardiac Catheterization screen.
To start the procedure simulation:
1. Push the
Start
softkey.
All the procedure steps are done manually.
2. Push the
Insert
softkey for the next step.
3. You simulate an increase and decrease of simulated pressure when you
push the
Increase Pressure
or
Decrease Pressure
softkeys.
4. Push the
Pullback
softkey to simulate the pressures when the catheter is
pulled back.
Push the
Stop
softkey to stop the procedure and go back to the IBP: Cardiac
Catheterization screen.