Vital Signs Simulator
ECG Function
23
glh019.bmp
Figure 8. Performance Wave Screen
To change the waveform:
1. Push
or
to highlight the
Waveform
value.
2. Push
.
3. Push
or
to highlight
Square
,
Sine
,
Triangle
, or
Pulse
in the list.
4. Push
to set the performance waveform and go back to the ECG:
Performance screen.
All waveforms have the same two variables: Rate and Amplitude.
To change the rate:
1. Push
or
to highlight the
Rate
value.
2. Push
.
3. Push
or
to highlight a value in the rate list. Table 12 lists the rates for
each type of wave.
Table 12. Performance Wave Rates
Performance Wave
Rates
Square and Triangle
0.125, 2.0, and 2.5 Hz
Sine
0.05, 0.5, 1, 2, 5, 10, 25, 30, 40, 50, 60, 100, and 150 Hz
Pulse
30 and 60 BPM
4. Push
to set the rate and go back to the ECG: Performance screen.
To change the amplitude:
1. Push
or
to highlight the
Amplitude
value.
2. Push
.
3. Push
or
to adjust the amplitude. Each push of a key increases or
decreases the amplitude 0.05 mV between 0.05 mV and 1.00 mV and
0.25 mV above 1 mV to 5.00 mV in the direction of the key pushed.
4. Push
to set the rate and go back to the ECG: Performance screen.