![Fluke Biomedical ProSim 8 Series User Manual Download Page 43](http://html1.mh-extra.com/html/fluke-biomedical/prosim-8-series/prosim-8-series_user-manual_2310745043.webp)
Vital Signs Simulator
Invasive Blood Pressure Simulation and Tests
31
the display and updates with real-time simulation data. The time until the
contraction ends is also shows on the display. See Figure 15.
glh027.bmp
Figure 15. Fetal ECG Intrauterine Pressure Screen
If the
Contraction
value is set to
Manual Start
, only one contraction cycle is
simulated.
The timer shows the time until the contraction ends and the subsequent
contraction starts.
Push the
Stop
softkey to stop contractions and go back to the Fetal ECG: IUP
screen.
Invasive Blood Pressure Simulation and Tests
The Product simulates blood pressure for Invasive blood pressure monitors. You
can set each blood pressure variable through the front-panel controls. The
Product also simulates Swan-Ganz and cardiac catheterization procedures.
Set the Invasive Blood Pressure Variables
The Product can simulate two invasive blood pressure transducers at one time.
The blood pressure variables are set separately for each channel. To set these
variables, push
to show the IBP screen in Figure 16 on the display.
glh009.bmp
Figure 16. Invasive Blood Pressure Screen
Summary of Contents for ProSim 8 Series