50
Adjust NIBP Pulse Envelope
To modify the shift of the blood pressure envelope in the Product with
firmware 2.00:
1. Push
to show the NIBP screen.
2. Push
or
to highlight
Envelope Shift
.
3. Push
.
4. Push
to shift the blood pressure envelope to the right, or push
to shift
the blood pressure envelope to the left.
The default value for a shift is zero, and the range is from -10 % to +10 %. In
Figure 33, the top graph shows a blood pressure envelope with a negative
shift, and the bottom graph shows a blood pressure envelope with a positive
shift.
Negative Shift
1.0
0.8
0.6
0.4
0.2
0
40
80
120
160
260
mmHg
Amplitude
Positive Shift
120/80/Envelope
1.0
0.8
0.6
0.4
0.2
0
40
80
120
160
260
mmHg
Amplitude
glh059.eps
Figure 33. NIBP Blood Pressure Envelope Shifts