40
UMPIOXR400V2-7-1EN, 2012-07-01
9 Outputs
9.4 Function Check
The function of the installed output can now be tested. Connect a multimeter to the installed output.
Test of an analog output:
Test of a binary output:
The outputs are now configured. For the use of the serial output see chapter 10. For the activation of the outputs
see chapter 11.
Enter a test value (in the example, the current output is tested). It has
to be within the selected output range. Press ENTER.
If the multimeter displays the entered value, the output works cor-
rectly.
Select
YES
to repeat the test,
NO
to return to
SYSTEM SETTINGS
.
Press ENTER.
Select
REED-RELAIS OFF
or
OPEN COLLECT.OFF
in the scroll list
OUTPUT TEST
to test the de-energized state of the output. Press
ENTER.
Measure the resistance at the output. The value has to be high ohmic.
Select
YES
. Press ENTER.
Select
REED-RELAIS ON
or
OPEN COLLECT.ON
in the scroll list
OUTPUT TEST
to test the energized state of the output. Press
ENTER.
Measure the resistance at the output. The value has to be low ohmic.
Select
YES
to repeat the test,
NO
to return to
SYSTEM SETTINGS
.
Press ENTER.
I 1 : O u t p u t T e s t
4 m A
I 1 = 4 . 0 m A
A g a i n ? n o > Y E S <
B 1 : O u t p u t T e s t
R e e d - R e l a i s O F F
B 1 = O F F
A g a i n ? n o > Y E S <
B 1 : O u t p u t T e s t
R e e d - R e l a i s O N
B 1 = O N
A g a i n ? n o > Y E S <