9 Outputs
9 Outputs
If the measured values are to be transmitted to another instrument, the outputs have to be configured.
The outputs have to be installed in the program branch
SPECIAL FUNCTION
(see section 9.1) and to be activated
in the program branch
OUTPUT OPTIONS
(see chapter 11). The installation of an output consists of three steps:
•
assigning a measuring channel (source channel) to the output
•
defining the physical quantity (source item) to be transmitted to the output by the source channel and the prop-
erties of the signal
•
defining the behavior of the output in case no valid measured values are available
9.1 Installation of an Output
Note!
The settings will be stored at the end of the installation dialog. If the installation dialog is inter-
rupted by key BRK, the setting will not be stored.
Select
SPECIAL FUNCTION\SYSTEM SETTINGS\PROC. OUTPUTS
.
Press ENTER.
Select the output to be installed. The scroll list contains all actually
available outputs.
A tick
after a list item indicates that this output has already been
installed. Press ENTER.
This display is indicated if the output has not been installed yet. Select
YES
. Press ENTER.
If the output is already installed, select
NO
to reconfigure it or
YES
to
return to the previous menu item and to select another output. Press
ENTER.
Select in the scroll list the measuring channel to be assigned as
source channel to the output. Press ENTER.
This display will not be indicated if the transmitter has only one mea-
suring channel.
Select the source item type to be transmitted to the output. The differ-
ent source item types and source items are given in Tab. 9.1. Press
ENTER.
S y s t e m s e t t i n g s
P r o c . o u t p u t s
I n s t a l l O u t p u t
C u r r e n t I 1
I 1 e n a b l e
n o > Y E S <
I 1 d i s a b l e
n o > Y E S <
S o u r c e c h a n .
C h a n n e l A :
S o u r c e i t e m
M e a s u r e d v a l u e s