UMPIOXR400V2-7-1EN, 2012-07-01
23
6 Installation of PIOX R400, R400B, R400C, R400A2
•
Prepare the sensor cable with the cable gland.
•
Cut the outer shield and brush it back over the compression part.
Fig. 6.11: Cable gland
•
Screw the gasket ring side of the basic part in the housing.
•
Insert the sensor cable in the housing.
•
Fix the cable gland by screwing the cap nut on the basic part.
•
Connect the leads to the terminals of the transmitter (see Fig. 6.12).
Fig. 6.12: Direct connection of sensor PIOX R400B, R400C, R400A2 with transmitter PIOX TR374B, TR374C, TR374A2
cap nut
compression part
basic part
equipotential
bonding
terminal
terminal strip KL1
70 mm
10 mm
outer
shield
sensor
cable
terminal
sensor cable
m
easu
ri
ng
ch
anne
l A
V+
yellow
V-
green
A+
brown
B-
white
m
easu
ri
n
g
cha
nne
l B
2V+
yellow
2V-
green
2A+
brown
2B-
white
terminal strip KL1:
transmitter
Important!
To ensure good high frequency shielding, it is
important to provide good electrical contact
between the outer shield and the cable gland
and hence to the housing of the transmitter.
Cut the outer shield and brush it back as shown
in the figure.
Attention!
To connect the sensor cable to the
transmitter, use only metallic cable
glands.
outer
shield
Preparation
of extension cable and
sensor cable:
equipotential
bonding