A Technical Data
A Technical Data
Transmitter
transmitter
PIOX
TR374
TR374A2
TR374B
TR374C
TR379
standard field device field device for
ATEX zone 2
field device for con-
nection of sensors in
zone 0/1
field device for con-
nection of sensors in
zone 1
19 " module
power supply
100...240 V/50...60 Hz or
20...32 V DC
power consump-
tion
< 15 W
measuring chan-
nels
1, option: 2
signal damping
0...100 s, adjustable
response time
1 s
material
aluminum, powder coated
aluminum
degree of protec-
tion
according to EN
60529
IP 65
IP 20
dimensions see
dimensional drawing
42TE x 3HE
(without back panel)
see dimensional
drawing
weight
2.8 kg
1.7 kg
installation
wall mounting, option: 2 " pipe mounting
19 " rack mounting
operating tem-
perature
-10...+60 °C
display
2 x 16 characters, dot matrix, backlit
menu language
English, German
data logger
(option)
SD card 2 GB
-
-
-
explosion protection
zone
-
2
outside of explosive
atmosphere
outside of explosive
atmosphere
-
marking -
II3G
Ex nA nC ic IIC T4 Gc
T
a
-40...+60 °C
0637
II(1)G
[Ex ia] IIB
T
a
-40...+60 °C
0637
II(2)G
[Ex ib] IIC
T
a
-40...+60 °C
-
certification
ATEX
-
-
IBExU06ATEX1075 X IBExU06ATEX1075 X -
certification
IECEx
-
-
IECEx IBE 10.0003X IECEx IBE 10.0003X -
type of protec-
tion
-
non sparking
-
-
-
intrinsic safety
parameters
U
m
= 250 V
intrinsically safe
power supply for
zone 0/1 sensors:
P
o
= 1.1 W
U
o
= 13.2 V
I
o
= 84 mA
C
o
= 5.1
μ
F
L
o
/R
o
= 32.8
μ
H/
Ω
RS485 (intrinsic
safety):
P
o
= 0.2 W
U
o
= 6 V
I
o
= 76 mA
C
o
= 999.3
μ
F
L
o
/R
o
= 3.37 mH/
Ω
U
m
= 250 V
intrinsically safe
power supply for
zone 1 sensors:
P
o
= 1.1 W
U
o
= 13.2 V
I
o
= 84 mA
C
o
= 0.24
μ
F
L
o
= 10 mH
RS485 (intrinsic
safety):
P
o
= 0.2 W
U
o
= 6 V
I
o
= 76 mA
C
o
= 39.3
μ
F
L
o
= 10 mH
remark intrinsically
safe
power supply for
zone 0/1
sensors
included
intrinsically safe
power supply for
zone 1
sensors
included