Chapter 8: Diagnostic and Repair Procedures
8-25
9.
Secure the generator with the eight securing screw.
10.
Connect the new generator power supply cable provided in the
replacement kit to the power supply board and to the
generator.
11.
Reconnect connector PX-1 to the power supply board.
12.
Replace the FT-70’s cover and power up the unit.
13.
To complete the installation procedure you must recalibrate the
x-ray threshold as explained in
Chapter 4: Configuring the
X-ray Threshold
.
However, because the x-ray generator takes approximately 5
hours to burn in after it is powered up, you must perform the
x-ray calibration procedure twice:
•
Calibrate the x-ray threshold now so that you can begin
using the FT-70, but note that the inspection accuracy may
be slightly below expected levels.
•
Calibrate the x-ray threshold again after 5 hours of use.
This will allow you to fine tune the threshold to
compensate for the changes in the source over the burn-in
period. This will return the inspection accuracy back to
optimum level.
Caution
If your FT-70 is equipped with a scintillation tube detector (STD),
do not recalibrate the STD.
Industrial Dynamics Company, Ltd. ©
Form
40256-0600
Summary of Contents for FT-70c
Page 1: ...FT 70c Service and Technical 40256 0600 Reference Manual for Can Inspection...
Page 14: ...Table of Contents XIII Industrial Dynamics Company Ltd Form 40256 0600...
Page 21: ...FT 70 Fill Level Inspector Service Manual 1 2 Industrial Dynamics Company Ltd 40256 0600...
Page 33: ...FT 70 Fill Level Inspector Service Manual 1 14 Industrial Dynamics Company Ltd 40256 0600...
Page 35: ...FT 70 Fill Level Inspector Service Manual 2 2 Industrial Dynamics Company Ltd 40256 0600...
Page 39: ...FT 70 Fill Level Inspector Service Manual 2 6 Industrial Dynamics Company Ltd 40256 0600...
Page 41: ...FT 70 Fill Level Inspector Service Manual 3 2 Industrial Dynamics Company Ltd 40256 0600...
Page 69: ...FT 70 Fill Level Inspector Service Manual 3 30 Industrial Dynamics Company Ltd 40256 0600...
Page 71: ...FT 70 Fill Level Inspector Service Manual 4 2 Industrial Dynamics Company Ltd 40256 0600...
Page 139: ...FT 70 Fill Level Inspector Service Manual 5 44 Industrial Dynamics Company Ltd 40256 0600...
Page 141: ...FT 70 Fill Level Inspector Service Manual 6 2 Industrial Dynamics Company Ltd 40256 0600...
Page 183: ...FT 70 Fill Level Inspector Service Manual 7 2 Industrial Dynamics Company Ltd 40256 0600...
Page 225: ...FT 70 Fill Level Inspector Service Manual 9 2 Industrial Dynamics Company Ltd 40256 0600...
Page 233: ...FT 70 Fill Level Inspector Service Manual 10 2 Industrial Dynamics Company Ltd 40256 0600...
Page 243: ...FT 70 Fill Level Inspector Service Manual 10 12 Industrial Dynamics Company Ltd 40256 0600...