Chapter 6: Calibration Procedures
6-11
If the reject is late, check to be sure the encoder coupling is not
slipping. Try a rejection with a thin layer of grease on the pad.
If the resulting mark in the grease starts in the upstream third
of the pad face and is centered, then performance is acceptable.
If there are still significant early/late problems, and you are
certain that you have a standard conveyor chain, contact the
Industrial Dynamics Customer Service.
23.
If you performed the previous step (step 22) at production
speed, then proceed to step 27. If you did not perform the
previous step (step 22) at production speed, then you must
proceed to step 24.
24.
With the conveyor running at production speed, press Key 5
until CHAIN VELOCITY appears. Note this value because you
need to enter it in the following step.
25.
Press Key 5 until MAXIMUM CONVEYOR SPEED appears
and enter the CHAIN VELOCITY value from the previous
step. The optimum encoder value for the current line speed is
now calculated including the Calibrate Encoder Value, Encoder
Prescaler Value, and Gamma Sample Cutoff Limit.
26.
While the conveyor is running at production speed, place a can
on the conveyor far enough upstream so that it does not slip on
the conveyor as it passes through the FT-70.
If the reject is late, check to be sure the encoder coupling is not
slipping. Try a rejection with a thin layer of grease on the pad.
If the resulting mark in the grease starts in the upstream third
of the pad face and is centered, then performance is acceptable.
If there are still significant early/late problems, and you are
certain that you have a standard conveyor chain, contact the
Industrial Dynamics Customer Service.
27.
To store the settings in system memory, close dip switch S1-2
and then press Key 5 until SAVE DEFAULT CONSTANTS
appears.
28.
Press the Arrow Key to execute. When COMPLETED appears,
open S1-2 to protect the saved settings.
29.
Press Key 5 until SYSTEM DISPLAY appears and press the
Arrow Key to select the Operate Mode. This returns the FT-70
to a secured operating mode and completes the calibration
procedure.
Industrial Dynamics Company, Ltd. ©
Form
40256-0600
Summary of Contents for FT-70c
Page 1: ...FT 70c Service and Technical 40256 0600 Reference Manual for Can Inspection...
Page 14: ...Table of Contents XIII Industrial Dynamics Company Ltd Form 40256 0600...
Page 21: ...FT 70 Fill Level Inspector Service Manual 1 2 Industrial Dynamics Company Ltd 40256 0600...
Page 33: ...FT 70 Fill Level Inspector Service Manual 1 14 Industrial Dynamics Company Ltd 40256 0600...
Page 35: ...FT 70 Fill Level Inspector Service Manual 2 2 Industrial Dynamics Company Ltd 40256 0600...
Page 39: ...FT 70 Fill Level Inspector Service Manual 2 6 Industrial Dynamics Company Ltd 40256 0600...
Page 41: ...FT 70 Fill Level Inspector Service Manual 3 2 Industrial Dynamics Company Ltd 40256 0600...
Page 69: ...FT 70 Fill Level Inspector Service Manual 3 30 Industrial Dynamics Company Ltd 40256 0600...
Page 71: ...FT 70 Fill Level Inspector Service Manual 4 2 Industrial Dynamics Company Ltd 40256 0600...
Page 139: ...FT 70 Fill Level Inspector Service Manual 5 44 Industrial Dynamics Company Ltd 40256 0600...
Page 141: ...FT 70 Fill Level Inspector Service Manual 6 2 Industrial Dynamics Company Ltd 40256 0600...
Page 183: ...FT 70 Fill Level Inspector Service Manual 7 2 Industrial Dynamics Company Ltd 40256 0600...
Page 225: ...FT 70 Fill Level Inspector Service Manual 9 2 Industrial Dynamics Company Ltd 40256 0600...
Page 233: ...FT 70 Fill Level Inspector Service Manual 10 2 Industrial Dynamics Company Ltd 40256 0600...
Page 243: ...FT 70 Fill Level Inspector Service Manual 10 12 Industrial Dynamics Company Ltd 40256 0600...