Control IDEC ramp and soak controller
n.
Control method: Proportional hot gas and resistance heat
o.
Refrigeration: Cascade scroll compressors 5hp and 3.5hp
p.
Heat: 2600 Watts
q.
Shelves: Quantity 2, adjustable on .595” centers plus bottom of chamber
r.
Power requirement- North America: 208-230 Volt, 3 phase, 31 amps, 40 amp circuit
required. Hubbell Locking Plug, Hubbell part number: CS8365C 50A rated, 250VAC,
3P4W. Requires Receptacle Hubbell part number: CS8369
s.
Power requirement- Europe/China 400 Volt, 3 phase, wye connected, 24 amps. 30 amp
fuses required and IEC 60309 (6H) 30 Amp 200/415V receptacle.
t.
Thermal room load:
Air cooled system: 36,860 BTU/Hr/11kW maximum during chamber warm up.
Water cooled system: 1706 BTU/0.5 kW (maximum)
u.
Drain: External with heated evaporator pan, powered from the cabinet.
v.
Water Supply for water cooled condenser models 4102-LC & 4105-LC
Tower water at
85F (29.5C) and
20 psi differential (1.38 Bar):
Water flow rate at maximum load = 7.0 gpm (26.5 liters/ min)
Water flow rate at average load = 4.5 gpm (15.0 liters/ min)
Chilled water at
50F (10C) and
20 psi differential (1.38 Bar):
Water flow rate at maximum load = 3.5 gpm (13.3 liters/ min)
Water flow rate at average load = 2.0 gpm (7.6 liters/ min)