4000
Ͳ
LC
Series
D4580880
Ͳ
01
Page
38
of
61
5 – “Common” : Electrically connected as a common connection for each of the
above dry contacts.
Terminals 7 (+) and 8 (-) are a 0-10VDC retransmit signal for the chamber control probe
temperature.
Linear Scaling is: 0 VDC = -100 °C, +10 VDC = +100 °C.
Terminals 9 (+) and 10 (-) are a 0-10VDC retransmit signal for the product probe
temperature.
Linear Scaling is: 0 VDC = -100 °C, +10 VDC = +100 °C.
Door open alarm – horn and strobe
In the event of a door opening while a profile or defrost cycle is running, there is a
horn and strobe light installed on the top of the unit that will activate and remote
contact terminals 4 & 5 will close.