Error and Warning Messages
60
T100S-HP User Guide
Table 13: Error and Warning Messages
FEEDBACK:ERR
Active wavelength monitoring
does not work properly.
Possible causes are:
Press
Enter
to restore the current
system settings. Internal wavelength
referencing remains unchanged.
•
No communication
between the T100S-HP
tunable laser and the
wavemeter.
•
Check the RS-232 serial
connection between the two
devices.
•
The wavemeter may not be
compatible with the SCPI
command set.
•
Please contact a EXFO
representative for a list of
compatible wavemeters.
•
The wavemeter remote
interface may not be
configured properly.
•
Configure the wavemeter serial
port with the following
parameters:
•
RS-232
ON
•
Baud rate
9600
•
Handshaking
None
•
No optical signal.
•
Press the
Enable
key to turn on
the laser output
•
Wavemeter malfunction.
DEFAULT
Laser Head
Overheating
The temperature of the room is
too high, the laser is disabled
Switch off the product and make
sure the temperature room is lower
than 30°C before starting again the
product.
Message
Possible Cause
Troubleshooting
Summary of Contents for T100S-HP Essential
Page 1: ...www EXFO com T100S HP High Power Tunable Laser User Guide T100SHP_UM_6 06v1 2...
Page 2: ......
Page 10: ...Important Safety Information for Your Product 10 T100S HP User Guide...
Page 20: ...Product Presentation 20 T100S HP User Guide...
Page 24: ...Installing and Connecting the Product 24 T100S HP User Guide...
Page 42: ...Enabling Active Wavelength Monitoring 42 T100S HP User Guide...
Page 50: ...Using the Auxiliary Inputs and Outputs 50 T100S HP User Guide...
Page 54: ...Verifying Performances 54 T100S HP User Guide...
Page 58: ...Performing Basic Maintenance Operations 58 T100S HP User Guide...
Page 64: ...List of Tables 64 T100S HP User Guide...
Page 66: ...List of Figures 66 T100S HP User Guide...
Page 70: ...Index 70 T100S HP User Guide...