8. Using the Auxiliary Inputs and Outputs
The auxiliary input and output connectors are located on the rear panel of the T100S-HP,
as described in section
Description of the Rear Panel, p. 18.
8.1
Implementing External Modulation
8.1.1
Implementing Low Frequency Intensity Modulation
Subject
You can implement low frequency intensity modulation of the output beam:
•
in the constant-current mode:
•
from DC to 8 MHz for sinusoidal signals
•
from DC to 1 MHz for TTL signals
•
in the APC (constant-power) mode:
•
from 30 kHz to 8 MHz for sinusoidal signals
•
from 30 kHz to 1 MHz for TTL signals
Modulation of the diode current is obtained by applying a signal anywhere between
0 and
5 volts
on the BNC connector labeled
LF Mod
.
Before Starting
Procedure
1.
Connect an external modulation source to the
LF Mod
. BNC connector, as shown in
•
The output impedance of your waveform generator must match the 50
Ω
input
impedance of the T100S-HP
LF Mod
. BNC connector.
Important
Before implementing external modulation via
LF Mod
, input connector, make
sure the system’s active cavity control is disabled (see section
Enabling Active Cavity Control, p. 35
).
Summary of Contents for T100S-HP Essential
Page 1: ...www EXFO com T100S HP High Power Tunable Laser User Guide T100SHP_UM_6 06v1 2...
Page 2: ......
Page 10: ...Important Safety Information for Your Product 10 T100S HP User Guide...
Page 20: ...Product Presentation 20 T100S HP User Guide...
Page 24: ...Installing and Connecting the Product 24 T100S HP User Guide...
Page 42: ...Enabling Active Wavelength Monitoring 42 T100S HP User Guide...
Page 50: ...Using the Auxiliary Inputs and Outputs 50 T100S HP User Guide...
Page 54: ...Verifying Performances 54 T100S HP User Guide...
Page 58: ...Performing Basic Maintenance Operations 58 T100S HP User Guide...
Page 64: ...List of Tables 64 T100S HP User Guide...
Page 66: ...List of Figures 66 T100S HP User Guide...
Page 70: ...Index 70 T100S HP User Guide...