T100S-HP User Guide
39
7. Enabling Active Wavelength Monitoring
When associated with an external optical wavelength meter, the T100S-HP tunable laser
can implement active wavelength monitoring to achieve high-precision wavelength
tuning (±2 pm, typically or within the wavemeter’s accuracy).
7.1
Setting Up the Product for Active Wavelength Monitoring
Subject
This section outlines the product setup and system requirements.
Before Starting
The wavemeter used must meet the following hardware and configuration requirements:
•
Serial port command set: based on the SCPI (Standard Commands for Programming
Instruments) syntax.
•
Serial interface configuration:
•
RS-232: ON
•
Baud rate: 9600
•
Handshaking: None
Depending on the wavemeter you are using, configuration can be achieved either via
soft-panel controls or by hardware setting via a back-panel dip-switch. For details on how
to configure your wavemeter, please refer to the user’s manual provided by the
manufacturer.
As an example, with a Burleigh-EXFO wavemeter, use a DB9 male-male straight-through
serial cable to connect the T100S-HP RS-232 C port to the wavemeter serial port.
Please contact your EXFO representative for a list of compatible wavemeters.
Procedure
1.
To setup the measurement, connect the T100S-HP output to the wavemeter via the
low output port of an optical coupler.
The remainder of the signal available on the coupler’s high-output is sent to the
component under test, as shown in the following figure.
Figure 6: Instrument Setup for Wavelength Monitoring
Wavemeter
I
o
Wavemeter
Rear Panel
Front Panel
Component
under test
Other Test
Equipment
Coupler
Summary of Contents for T100S-HP Essential
Page 1: ...www EXFO com T100S HP High Power Tunable Laser User Guide T100SHP_UM_6 06v1 2...
Page 2: ......
Page 10: ...Important Safety Information for Your Product 10 T100S HP User Guide...
Page 20: ...Product Presentation 20 T100S HP User Guide...
Page 24: ...Installing and Connecting the Product 24 T100S HP User Guide...
Page 42: ...Enabling Active Wavelength Monitoring 42 T100S HP User Guide...
Page 50: ...Using the Auxiliary Inputs and Outputs 50 T100S HP User Guide...
Page 54: ...Verifying Performances 54 T100S HP User Guide...
Page 58: ...Performing Basic Maintenance Operations 58 T100S HP User Guide...
Page 64: ...List of Tables 64 T100S HP User Guide...
Page 66: ...List of Figures 66 T100S HP User Guide...
Page 70: ...Index 70 T100S HP User Guide...