8 Maintenance
Instructions for use
•
MERGITE!
73/84
Preparing the container with SETUP CLEAN (1:5)
WARNING
STRONGLY ALKALINE CLEANING SOLUTION
EUROIMMUN SETUP CLEAN is corrosive when in contact with
skin or eyes.
>
Wear protective gloves, gown and goggles.
Prepare the working solution for the container as follows:
500 ml EUROIMMUN SETUP CLEAN
2000 ml distilled water
Observe the instruction for use (ZG_0009_A_UK_BXX) when
preparing the working solution.
The working solution SETUP CLEAN (1:5) can be used
immediately.
8.4 Measures for storage or transport
Carry out the following steps before you store or transport the device.
a) Empty the liquid system.
Chapter 7.3 Draining MERGITE!, p. 62
b)
Clean/disinfect and empty all liquid containers.
Chapter 8.3.2 Cleaning/disinfection of the container, p. 70
c)
Clean the device and the accessories.
Chapter 8.3.3 Cleaning the device and accessories, p. 71
d)
Disconnect the device from the power supply.
e)
Package the device in the original packaging.
Observe the storage and transport conditions
(
Chapter 10.2 Environmental conditions, p. 77