8 Maintenance
68/84
Instructions for use
•
MERGITE!
4) Exchange the containers with distilled water (In 1) and wash
buffer (In 2) for containers with SETUP CLEAN (1:5).
Have a paper towel at the ready on which to place the container
cap and to mop up any dripping liquid residues.
5) Press
START
.
The system is filled with SETUP CLEAN (1:5) and the
progress indicator for the exposure time appears.
After the end of the exposure time, the prompt appears to
reconnect the containers with distilled water and wash buffer.
Fig. 46 Connect container with SETUP CLEAN (1:5)
Fig. 47 Display of the exposure time