8 Maintenance
Instructions for use
•
MERGITE!
65/84
8.1 Maintenance overview
8.1.1 Maintenance at the start of work
When switching on the device, the liquid system is automatically
rinsed with wash buffer.
8.1.2 Maintenance at the end of work
When switching off the device, the liquid system is automatically
rinsed and filled with distilled water.
Disinfection of the wash basin
Cleaning as needed
8.1.3 Weekly maintenance
Disinfection of the wash buffer container
Chapter 8.3.2 Cleaning/disinfection of the container, p. 70
Replacement of wash buffer
Cleaning the surfaces of the device and accessories
Chapter 8.3.3 Cleaning the device
8.1.4 Monthly maintenance
Cleaning of the liquid system
Chapter 8.3.1 Cleaning of the liquid system, p. 67
Cleaning and disinfection of the containers for distilled water,
wash buffer and waste liquid
Chapter 8.3.2 Cleaning/disinfection of the container, p. 70
Cleaning the surfaces of the device and accessories
Chapter 8.3.3 Cleaning the device and accessories, p. 71
8.1.5 Annual maintenance
The annual maintenance on the device is carried out by the
Technical Customer Support of the EUROIMMUN AG and includes,
among other things, the inspection of wearing parts.