Position the calibration gauge in the posiblock holder.
Press
to tighten the clamps.
Press
to start the calibration cycle.
The gauge is positioned on the feeler level.
A message is displayed: "Are the feelers aligned in the plastic pattern hole?"
Check that the feelers are correctly aligned.
Press
if the feelers are correctly aligned.
Press
to loosen the clamps and remove the plastic pattern, then
to return to the work screen.
The calibration cycle is complete.
Press
if the feelers are not aligned.
A message is displayed indicating that calibration by a technician is required: contact your technical
department.