Controller Operating Environmental Conditions
0 to 33 Celsius
5 to 90% RH Non-Condensing
Microscope Operating Environmental Conditions
0 to 42 Celsius
5 to 95% RH Non-Condensing
2.5 Unpacking and Checking the IncuCyte™
The installation of a new IncuCyte™ ZOOM system will be done by trained Essen BioScience field
service personnel. The process is described in the sections below and should be followed if you
ever move the system to a new location.
The IncuCyte™ ZOOM System is shipped in two boxes, one containing the automated microscope
and the other containing the controller.
NOTE:
The IncuCyte™ ZOOM consists of 2 components; the controller and the microscope. The
controller remains outside the cell culture incubator, while the microscope is place inside the
incubator and holds the cells for monitoring. These two components are shipped in separate
boxes.
To unpack the microscope, start by setting the box with the correct side up on the floor or on a
sturdy bench top. Start removing materials from the top as shown in the exploded diagram
(Figure 1).