Environnement
S.A
AF22e
Duplication prohibited
JUNE 2016
4–14
MAINTENANCE SHEET
MONITOR Serial No.:
OPERATION SHEET: 4.3.5
Scope: UV lamp replacement
PAGE: 1/3
Frequency: 2 years
WARNING: DO NOT put your fingers on the lamp glass.
Dismantling the UV source assembly:
Turn off the analyzer and unplug the power cable.
Remove the cover screw located on the analyzer rear panel. Remove the protective cover.
Unplug the UV source supply board (3) connector (2) to disconnect the UV source assembly (1).
Slightly pull up the measurement c UV source assembly to remove from the foam block (4), and
loosen the two UV source holder screws (5). Then, handle the lower part (6) of the UV source assembly to pull
it downwards as arrows (7) indicate.
Remove the UV source assembly (3) from its holder (8).
(1) UV source supply board, (2) UV source connector, (3) UV source assembly, (4) foam block, (5) UV source
holder screws, (6) lower part of UV source assembly, (7) arrows, (8) UV source holder.
Figure 4–7 – UV source assembly dismantling
Tools and parts required
Screwdriver, 5.0 x 75 mm
UV lamp, Ref. : D01-1297
Summary of Contents for AF22e
Page 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Page 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Page 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Page 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Page 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Page 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Page 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Page 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...