Environnement
S.A
AF22e
Duplication prohibited
JUNE 2016
4–8
MAINTENANCE SHEET
MONITOR Serial No.:
OPERATION SHEET: 4.3.3
Scope: Internal zero filter replacement
PAGE: 1/1
Frequency: 6 months
Proceed in the following order:
Turn off the analyzer and unplug the power cable.
Remove the cover screw located on the analyzer rear panel. Remove the protective cover.
Disconnect the hydrocarbon scrubber tube (1) and the solenoid valve tube (2).
Remove the internal zero filters (3) and replace with new ones.
Reconnect the hydrocarbon scrubber tube (1) and the solenoid valve tube (2).
Plug in the power cable and power on the analyzer.
Click-on the alarm icon displayed in the bottom bar of any screen to view active alarms. Then,
click-on the internal zero filter alarm and set the internal zero filter countdown to 180 days.
Wait at least one hour for the measurement to stabilize, then launch a new zero-reference cycle.
(1) hydrocarbon scrubber tube, (2) solenoid valve tube, (3) internal zero filters.
Figure 4–2 – Internal zero filter replacement
Dates
Tools and parts required:
Internal zero air filter. Ref.: F05-0128-C-SAV
Summary of Contents for AF22e
Page 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Page 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Page 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Page 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Page 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Page 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Page 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Page 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...