Environnement
S.A
AF22e
Duplication prohibited
APRIL 2016
3–36
« Span gas value »
field: span gas cylinder concentration used to perform calibration check. This
concentration is the reference value to be reached.
« Calib. ON/OFF »
field: this field is used when the analyzer can measure many gases. It allows to
perform a gas-by-gas calibration with no effect on the other gases. It invalidates or not the auto-
calibration for the considered gas.
« Response time »
field: shows response time value. Refer to Chapter 2 of this manual for more details.
« Filter autonomy (d) »
field: shows the new zero air filter autonomy (day number). The analyzer
decrements each day from this value (refer to synoptic display of the standard function screens). A
control triggers when it lasts 120-day autonomy, and an alarm is triggers when it lasts one-day
autonomy.
« Kicker autonomy »
field (d): shows the new carbon kicker filter autonomy (day number). The analyzer
decrements each day from this value (refer to synoptic display of the standard function screens). A
control is triggers when it lasts 120-day autonomy, and an alarm is triggers when it lasts one-day
autonomy.
« Init. Counter OFF/ON »
field: reset filter autonomy to the programmed value after a filter change.
« Dilution OFF/ON »
field: used to activate the dilution function and to set the dilution coefficient.
3.4.2.12.
Automatic linearization of the flow
This screen is accessed by touching/clicking-on
. It allows to linearize the flow value as function of
the corresponding pump set point.
Before linearization launching, the user must connect a flowmeter (0-100 nl/h) to the suitable fluid inlet.
Then, he clicks-on/touches the corresponding fluid inlet button
, or
, or
, and he inputs in
the «
Nominal
» field the read value on the flowmeter. Then, he launches linearization by
touching/clicking on
. Then, the analyzer will be in a maintenance mode for around 5 minutes.
REMARK: the in-factory pump setting is 100% by-default.
Summary of Contents for AF22e
Page 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Page 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Page 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Page 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Page 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Page 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Page 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Page 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...