MAINTENANCE SHEET
MONITOR Serial No.:
OPERATION SHEET: 4.3.2
Scope: Optical, electric and fluid parameter check
PAGE: 1/2
Periodicity :
1 month
Flow rate check :
•
Connect a flow meter on the AF22e sample inlet.
•
Make sure the flow rate value indicated on the flow meter is 20 ± 5 Nl/h.
Date
Tools required
•
Flow meter, 100 liters/hour
Sample inlet
Flow meter