Duplication prohibited
AF22e
Environnement
S.A
JUNE 2016
4–7
MAINTENANCE SHEET
MONITOR Serial No.:
OPERATION SHEET : 4.3.2
Scope : Optical, electric and fluid parameter check
PAGE : 2/2
Periodicity :
1 month
MUX signals check:
Switch the analyzer to zero air.
Compare the values given on the
« Diagnostic functions (alarm, input/output, mux…)»
screen
with the limits listed in the table below.
Signals Min.
values
Typical
values
Max values
UV ref.
250 mV
N. A.
N. A.
HV PM
1600 mV
2800 mV
4000 mV
Internal T.
5 °C
30 °C
50 °C
Measurement chamber T°C
48°C
50 °C
52 °C
Gas pressure
600 hPa
990 hPa
1200 hPa
Pump pressure
200 hPa
300 hPa
600 hPa
+4V supply
3950 mV
4000 mV
4050 mV
+3.3 V supply
3.10 V
3.30 V
3.50 V
+12V supply
11.50 V
12 V
12.60 V
+24 V supply
23.0 V
24.0 V
25.0 V
+5 V supply
4.70 V
5 V
5.30 V
5 V supply
4.80 V
5 V
5.20 V
+24V current
- 0.1 A
0.5 A to 3 A
> 6 A
Date
Tools required
Flow meter, 100 liters/hour
Summary of Contents for AF22e
Page 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Page 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Page 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Page 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Page 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Page 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Page 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Page 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...