Alarm message
(Min./Max. values)
Possible cause(s)
Possible corrective action(s)
–
If pump pressure is abnormally high:
•
Pressure sensor is faulty.
•
A fluid tube is unplugged.
•
Replace pressure sensor.
•
Check the fluid tubes.
Flow Fault
Min. value: 5 l/h
Max. value: 90 l/h
–
Pump cable is disconnected.
–
A fluid tube is unplugged.
–
Interconnection board pump
command is inactive.
–
Pressure sensor’s inlet ports of
pump and gas are disconnected.
–
Connect pump cable to
interconnection board.
–
Check the fluid tubes.
–
Check that interconnection board
pump command is active and
controller board operates correctly.
–
Check pump and gas pressure sensor
connections.
No cham. Temp
Min. value: N.A.
Max. value: N.A.
–
Chamber temperature sensor cable
is disconnected.
–
Connect chamber temperature sensor
cable.
Cham. T open
Min. value: N.A.
Max. value:
> 70°C
–
Chamber temperature sensor is
connected but connection is faulty
(as an open circuit).
–
Replace temperature sensor: analyzer
exits alarm mode after 60 sec.
Cham. T shortcut
Min. value: < 5°C
Max. value: N.A.
–
Chamber temperature sensor is
connected, but connection is faulty
(as a closed circuit).
–
Replace chamber temperature
sensor: analyzer exits alarm mode
after 60 sec.
Chamber temperature
Min. value: 48°C
Max. value: 52°C
–
Controller board configuration is
faulty.
–
Heating laminar is not connected.
–
Heating laminar is faulty.
–
Replace controller board with a
calibrated controller board.
–
Connect heating laminar.
–
Replace heating laminar.
No internal Temp
Min. value: absent
Max. value: absent
–
Internal temperature sensor cable is
not connected.
–
Connect internal temperature sensor
cable.
Summary of Contents for AF22e
Page 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Page 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Page 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Page 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Page 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Page 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Page 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Page 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...