Duplication prohibited
AF22e
Environnement
S.A
APRIL 2016
4–5
MAINTENANCE SHEET
MONITOR Serial No.:
OPERATION SHEET: 4.3.1
Scope: Sample inlet dust filter replacement
PAGE: 1/1
Frequency: 2 to 4 weeks
according to dust amount
Proceed in the following order:
−
Turn off the analyzer power switch and disconnect the sample inlet.
−
Unlock the filter holder cover (1) by pulling the spring (2) upwards.
−
Remove the worn filter (3).
−
Place the new filter (4) on the filter holder without the separating blue sheet.
−
Make sure that the filter correctly covers the filter-holder surface and that it is correctly placed flat
on top of it.
−
Reposition the filter holder cover, and lock it by lowering the spring (2).
−
Connect the sample inlet and power on the analyzer.
(1) filter holder cover, (2) spring, (3) worn filter, (4) new filter.
Figure 4–1 – Sample inlet dust filter replacement
Date
Tools and parts required:
•
Tweezers
•
Sample inlet filter (Teflon filter - porosity 5 µm -
∅
47 mm) - Ref.: F05-11-842
Summary of Contents for AF22e
Page 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Page 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Page 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Page 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Page 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Page 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Page 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Page 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...