Duplicationh prohibited
AF22e
Environnement
S.A
3–45
APRIL 2016
−
Dilution air (zero air):
Zero air generator or cylinder of reconstituted air.
Zero air should not contain more than 0.5 ppb of SO
2
and be free of contaminants that might be
measured by the analyzer.
−
Span gas (SO
2
):
Certified SO
2
cylinder, concentration 2% in air allowing six-range generation between 15 % and 90
% of the full scale of the measurement range used. Carrier gas may be nitrogen if the system allows
a higher dilution ratio than 600:1.
3.5.4.3. Procedure
−
Turn on the analyzer at least 6 hours before performing calibration.
−
Touch the
key to display
« Automatic cycle configuration»
screen and program analyzer as
follows (advised configuration) :
−
Touch the
key to display
« Advanced analyzer configuration »
screen:
Response time: set
Auto
–
Configure
SO
2
offset to 10 ppb at least, or 5 % of the full scale of the measurement range used.
Summary of Contents for AF22e
Page 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Page 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Page 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Page 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Page 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Page 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Page 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Page 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...