26
input lines
IEC 61000-4-11
< 5 % U
T
(> 95 % dip in U
T
) for 1
cycle
70 % U
T
(30 % dip in U
T
) for 25
cycles
< 5 % U
T
(> 95 % dip in U
T
) for 5 s
< 5 % U
T
(> 95 % dip in U
T
) for 1
cycle
70 % U
T
(30 % dip in U
T
) for 25
cycles
< 5 % U
T
(> 95 % dip in U
T
) for 5 s
it is recommended that the NanoVi
®
Wellness be powered from
an uninterruptible supply or a battery.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical commercial or hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer‘s declaration
- electromagnetic immunity
The NanoVi
®
Wellness is intended for use in the electromagnetic environment specified below. The customer or the user of the NanoVi
®
Wellness should
assure that it is used in such an environment.
Immunity test
lEC 60601 test
level
Compliance level
Electromagnetic environment - guidance
Conducted FR
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
10 V/m
80 MHz to 2,7 GHz
3 V
10 V/m
Portable and mobile RF communications equipment should be
used no closer to any part of the NanoVi
®
Wellness, including
cables, that the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance
𝑑 =
3,5
3
√𝑃
𝑑 =
3,5
10
√𝑃
80 MHz to 800 MHz
𝑑 =
7
10
√𝑃
800 MHz to 2,5 GHz
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).