25
16
Electromagnetic compatibility
Electromagnetic emissions
Guidance and manufacturer‘s declaration
- electromagnetic emissions
The NanoVi
®
Wellness is intended for use in the electromagnetic environment specified below. The customer or the user of the NanoVi
®
Wellness should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 2
The NanoVi
®
Wellness must emit electromagnetic energy in order to perform its intended
function. Nearby electronic equipment may be affected.
RF emissions
CISPR 11
Class B
The NanoVi
®
Wellness is suitable for use in all establishments, including domestic and
medical establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electromagnetic immunity
Guidance and manufacturer‘s declaration
- electromagnetic immunity
The NanoVi
®
Wellness is intended for use in the electromagnetic environment specified below. The customer or the user of the NanoVi
®
Wellness should
assure that it is used in such an environment.
Immunity test
lEC 60601 test
level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV air
± 4 kV air
± 8 kV air
± 15 kV air
± 8 kV contact
± 2 kV air
± 4 kV air
± 8 kV air
± 15 kV air
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 2 kV for power supply
lines
Mains power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 0,5, 1,0 kV line(s) to
line(s)
± 0,5, 1,0, 2 kV line(s) to
earth
± 0,5, 1,0 kV line(s) to
line(s)
± 0,5, 1,0, 2 kV line(s) to
earth
Mains power quality should be that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
< 5 % U
T
(> 95 % dip in U
T
) for 0,5
cycle
< 5 % U
T
(> 95 % dip in U
T
) for 0,5
cycle
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the NanoVi
®
Wellness
requires continued operation during power mains interruptions,