Technical data
Magphant
36
Hauser
15.4 Output
Output signal
Current output 4 to 20 mA
Signal mode
Active
Load
0 to 750 Ω
Bidirectional flow
measurement
The measuring device is able to measure in both flow directions, i.e. it supports
bidirectional measurement. The current output is always positive. The relay is
activated in both flow directions.
[mA]
20.5
20
[m/s]
[m/s]
2
2
1
3
A0040249
1
Forward flow
2
Scalable full scale values
3
Reverse flow
Relay output
Floating changeover contact
• 60 V AC/0.4 A
• 75 V DC/0.5 A
15.5 Power supply
Terminal assignment
Supply voltage
24 V
DC
(20 to 30 V
DC
)
The power unit must be tested to ensure it meets safety requirements (e.g. PELV, SELV).
Power consumption
<2.5 W
Electrical connection
Potential equalization
Cable specification