Installation
Hauser
15
90
°
2
1
A0040155
2
Position of electrode axis
1
Axis of electrodes
2
Axis of cable glands
Inlet and outlet runs
If possible, install the sensor upstream from fittings such as valves, T-pieces or elbows.
Observe the following inlet and outlet runs to comply with accuracy specifications:
5…10 × DN
5…10 × DN
A0041744
For information on the dimensions and installed lengths of the device → 38
Mounting conditions for welding socket
Installation in steel pipes
The measuring device is mounted in steel pipes using the welding socket supplied. Two
different types of welding socket are available, depending on the nominal diameter:
• Welding socket for pipes DN 25
• Welding socket for pipes ≥DN 40
NOTICE
Damage to the measuring device
‣
Only weld the welding socket when the measuring device is not installed.
‣
Pipe DN 25: Weld the welding socket at right angles to the axis of the piping.