Installation
Magphant
18
Hauser
Installation conditions for T-fitting
For pipe sizes DN 15 to 50, the measuring device is mounted in plastic pipes using a
standard T-fitting. The T-fitting can be purchased from the Georg Fischer company. PVC,
PP and PE T-fittings are available.
Only use the Magphant version for device installation in plastic pipes (order code for
"Process connection", option 5 "Adapter, plastic pipe, 316L, NBR").
236 (9.29)
DN
1
A0040255
6
Installation conditions for T-fitting. Engineering unit mm (in)
1
Standard T-fitting
6.1.2
Environment and process requirements
Ambient temperature range
–20 to +60 °C (–4 to +140 °F)
If operating outdoors:
• Install the measuring device in a shady location.
• Avoid direct sunlight, particularly in warm climatic regions.
• Avoid direct exposure to weather conditions.
6.2
Mounting the measuring device
6.2.1
Required tool
For sensing element
For the gland of the sensing element: use the appropriate installation tool.
6.2.2
Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove any protective covers or protective caps present from the sensor.
3. Remove stick-on label on the electronics compartment cover.
6.2.3
Mounting the sensor