Maintenance
Liquisys M CCM223/253
88
Hauser
8.1.5
Maintenance pH connecting lines and junction boxes (EP)
Check the cables and connections for moisture. Moisture is indicated by a sensor slope that
is too small. If no more display is possible or if the display is fixed at pH 7, please check the
following components:
• Sensor head
• Sensor connector
• pH measuring cable
• Junction box, if fitted
• Extension cable
NOTICE
Faulty measurement due to moisture in the measuring cable
‣
If there is moisture in the measuring cable, the cable must be replaced!
A shunt in the cable of > 20 M
can no longer be measured with normal multimeters but is
damaging for the pH measurement. A reliable test can be carried out with a usual
commercial insulation meter:
• Make sure to disconnect the pH measuring cable from the sensor and device!
• If you are using a junction box, check the infeed and outfeed measuring cable separately.
• Check the cable with 1000 V DC (at least with 500 V DC) testing voltage.
• If the cable is intact, the insulation resistance > 100 G
.
• If the cable is defective (moist), there is flashover.
The cable must be replaced.
The sensor head and junction box can be cleaned and dried with a hot air dryer.