Diagnostics and troubleshooting
Liquisys M CCM223/253
80
Hauser
Chlorine value too low
Measuring chamber not closed
Refill and carefully screw tight.
Electrolyte CCY14-F or CCY24-F
Air cushion outside in front of
membrane
Remove air bubble, possibly choose
better installation position.
Air cushion under membrane
Refill and carefully screw tight bubble-
free.
Electrolyte CCY14-F or CCY24-F
Chlorine value wrong / no
calibration possible; zero
current too large
Wrong polarisation voltage
Measure polarisation voltage, replace
MKC1 module if required.
DVM (+) at S or 90, DVM (-) at 91
CCS140/141: -20 mV
CCS240/241: +120 mV
Wrong sensor type selected
Check sensor selection.
pH/mV measurement:
Measuring chain zero
point not adjustable
Reference system poisoned
Test with new sensor.
pH/mV sensor
Membrane blocked
Clean / grind membranes (not with
TEFLON membranes).
HCl 3%, warding file (only file in one
direction); new sensor
Measuring line broken
Short-circuit pH input and connect to PM
-> reading pH 7.
Asymmetrical voltage of sensor too
high
Clean membranes or test with different
sensor.
HCl 3%, file (only file in one direction);
new sensor.
Potential matching (PM) of Liquisys
M -> medium incorrect
CCM223/253 always connect
symmetrically to PM!
Connection diagram
No or slow change of
readings
Sensor dirty
Clean sensor.
Cleaning agent
Sensor aged
Replace sensor.
New sensor
Sensor defective (reference outgoing
lead)
Replace sensor.
New sensor
Measuring chain slope
not adjustable /
Slope too low
Connection not at high impedance
(humidity, dirt)
Test cable, connector and junction boxes. pH simulator, insulation tester
Instrument input defective
pH measurement directly on the
instrument.
pH simulator
Sensor aged
Replace sensor.
pH sensor
Measuring chain slope
not adjustable /
No slope
Hairline crack in glass membrane
Replace sensor.
pH sensor.
Connection not at high impedance
Test cable, connector and junction boxes. pH simulator, insulation tester
Permanent, incorrect
reading
Sensor not immersed or protective
cap not removed
Check installation position, remove
protective cap.
Air cushion in assembly
Check assembly and installation
position.
Ground fault at or in the instrument Test measurement in insulated vessel,
possibly with buffer solution.
Plastic vessel, buffer solutions
Hairline crack in glass membrane
Replace sensor.
pH sensor
Instrument in impermissible
operating state (no response to key
actuation)
Switch instrument off and on.
Possibly EMC problem: If repeated, check
grounding and wire routing
Incorrect temperature
reading
Incorrect sensor connection
Check connections.
Connection diagrams
Sensor or measuring cable defective Test sensor and cable.
Ohmmeter
pH value in process
incorrect
Flow rate too high
Reduce flow rate or measure in a bypass.
Potential in medium
Possibly ground with / at PM pin
(connect PA/PM to PE).
Problem mainly occurs in plastic lines
Sensor covered in dirt or deposits
Clean sensor.
For highly polluted media: Use spray
cleaning
Error
Possible cause
Tests and / or remedial measures
Equipment, spare parts, personnel
Summary of Contents for Liquisys M CCM223
Page 109: ...Liquisys M CCM223 253 Technical data Endress Hauser 109...
Page 116: ...Appendix Liquisys M CCM223 253 116 Endress Hauser a0002639 en...
Page 120: ...Appendix Liquisys M CCM223 253 120 Endress Hauser...
Page 123: ...Liquisys M CCM223 253 Endress Hauser 123...
Page 124: ...www addresses endress com 71212626...