Liquisys M CCM223/253
Maintenance
Hauser
85
8
Maintenance
WARNING
!
Process pressure and temperature, contamination, electrical voltage
Risk of serious or fatal injury
‣
If the sensor has to be removed during maintenance work, avoid hazards posed by
pressure, temperature and contamination.
‣
De-energize the device before opening it.
‣
Power can be supplied to switching contacts from separate circuits. De-energize these
circuits before working on the terminals.
Take all the necessary measures in time to guarantee the operational safety and reliability
of the entire measuring system.
Maintenance work at the transmitter comprises:
• Calibration (see "Calibration" section)
• Cleaning of assembly and sensor
• Cable and connection check
When performing any work on the device, bear in mind any potential impact this may have
on the process control system or on the process itself.
NOTICE
Electrostatic discharge (ESD)
Risk of damage to electronic components
‣
Take personal protective measures to avoid ESD, such as discharging beforehand at PE or
permanent grounding with a wrist strap.
‣
For your own safety, use only genuine spare parts. With genuine spare parts, the
function, accuracy and reliability are also guaranteed after repair.
8.1
Maintenance of the entire measuring point
8.1.1
Cleaning the transmitter
Clean the front of the housing with usual commercial cleaning agents.
In accordance with DIN 42 115, the front is resistant to:
• Ethanol (short periods)
• Diluted acids (max. 2% HCl)
• Diluted bases (max. 3% NaOH)
• Soap-based household cleaners
NOTICE
Prohibited cleaning agents
Damage to the housing surface or housing seal
‣
For cleaning purposes, never use concentrated mineral acids or bases.
‣
Never use organic cleaners such as benzyl alcohol, methanol, methylene chloride, xylene
or concentrated glycerol cleaner.
‣
Never use high-pressure steam for cleaning purposes.
Summary of Contents for Liquisys M CCM223
Page 109: ...Liquisys M CCM223 253 Technical data Endress Hauser 109...
Page 116: ...Appendix Liquisys M CCM223 253 116 Endress Hauser a0002639 en...
Page 120: ...Appendix Liquisys M CCM223 253 120 Endress Hauser...
Page 123: ...Liquisys M CCM223 253 Endress Hauser 123...
Page 124: ...www addresses endress com 71212626...