14. In A6, select digital input 1.
Press
F
to confirm.
Hold = external hold
Clean = Cleaning
trigger
15. If you need to stabilise the display, enter the
required damping factor in A7.
Press
F
to confirm.
The display returns to the initial display of "Setup
1"
1
1 to 60
16. Press PLUS and MINUS simultaneously to switch
to the measurement mode.
User input
Setting range
(Factory settings,
bold)
Display
A6
Digital 1
Hold
A7
Damping
1