Commissioning
Liquisys M CCM223/253
68
Hauser
R279
Enter number of
cleaning cycles
without cleaning
agent (economy
function)
0
0 to 9
After cleaning with cleaner, up to 9
cleaning sessions can be carried out with
water only until the next cleaning session
with cleaner takes place.
R2 (8)
Configure 3-point
step controller for
Cl
2
/ClO
2
3PSt
Only with relays 3 and 4
By confirming with ENTER, another relay
function already switched on is switched
off
R281
Switch function of
R2 (8) off or on
Off
On
R282
Enter set point
500 ppb
0 to 20000 ppb
0.5 ppm
0 to 20 ppm
0.5 (0.1) mg/l
0 to 20 (5) mg/l
The set point is the value which the
controller has to hold. The controller will
restore this value if there is a deviation up
or down.
R283
Enter control gain K
p
1.00
0.01 to 20.00
See "P(ID) controller" section.
R284
Enter integral action
time T
n
(0.0 = no I-
component)
0.0 min
0.0 to 999.9 min
See "P(ID) controller" section.
R285
Enter minimum
switch-on time T
ON
0.3 s
0.1 to 5.0 s
R286
Enter motor run
time
60 s
10 to 999 s
Actuator motor run time from "fully
closed" state to "fully open" state.
R287
Enter neutral zone
10 %
0 to 40 %
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
R279
EconomyCl
0
R2
Function
3PSt
R281
Function
Off
R282
mg/l
Setpoint
0.50
R283
Kp
1.00
R284
min
Time Tn
0.0
R285
s
Min.PTime
0.3
R286
s
MotorTime
60
R287
%
NeutrZone
10
Summary of Contents for Liquisys M CCM223
Page 109: ...Liquisys M CCM223 253 Technical data Endress Hauser 109...
Page 116: ...Appendix Liquisys M CCM223 253 116 Endress Hauser a0002639 en...
Page 120: ...Appendix Liquisys M CCM223 253 120 Endress Hauser...
Page 123: ...Liquisys M CCM223 253 Endress Hauser 123...
Page 124: ...www addresses endress com 71212626...