Liquisys M CCM223/253
Diagnostics and troubleshooting
Hauser
77
E038
Cl sensor signal outside
permitted range during
slope calibration
• Service sensor (acc. to sensor
instructions)
• Check connections.
• Check DPD measuring instrument.
• Do not use organic chlorination
agent.
No
No
—
—
2
80
80
E041
Calibration parameter
calculation aborted
Repeat calibration and replace buffer
solution; if necessary, replace sensor
and check instrument and measuring
cable.
No
No
—
—
2
80
80
E042
Distance between buffer
value and zero point (pH7)
too small (single-point
calibration)
For slope calibration, use a buffer
solution which has at least a distance of
pH = 2 to the electrode zero point.
No
No
—
—
2
80
80
E043
Distance between buffer 1
and buffer 2 calibration
values too small (two-
point calibration)
Use buffer solutions which differ by at
least
pH = 2.
No
No
—
—
2
80
80
E044
Stability factor not met
during calibration
Repeat calibration and replace buffer
solution; if necessary, replace sensor
and check instrument and measuring
cable using simulator.
No
No
—
—
2
80
80
E045
Calibration aborted
No
No
—
—
2
80
80
E055
Lower Cl/ClO
2
measuring
range exceeded
Check measurement and connections; if
necessary, check instrument and
measuring cable using simulator.
Yes
No
No
80
E056
Lower pH/mV measuring
range exceeded
Yes
No
No
80
E057
Upper Cl/ClO
2
measuring
range exceeded
Yes
No
No
80
E058
Upper pH/mV measuring
range exceeded
Yes
No
No
80
E059
Lower temperature
measuring range exceeded
Yes
No
No
80
E061
Upper temperature
measuring range exceeded
Yes
No
No
80
E063
Lower current output
range1 exceeded
Check measured value and current
assignment.
Yes
No
No
80
80
E064
Upper current output
range1 exceeded
Yes
No
No
80
80
E065
Lower current output
range2 exceeded
Yes
No
No
80
80
E066
Upper current output
range2 exceeded
Yes
No
No
80
80
E067
Alarm threshold limit
contactor 1 exceeded
Check configuration.
Yes
No
No
80
80
E068
Alarm threshold limit
contactor 2 exceeded
Yes
No
No
80
80
E069
Alarm threshold limit
contactor 3 exceeded
Yes
No
No
80
80
E070
Alarm threshold limit
contactor 4 exceeded
Yes
No
No
80
80
E080
Current output 1 range
too small
Increase range in "Current outputs"
menu.
No
No
—
—
2
80
80
E081
Current output 2 range
too small
No
No
—
—
2
80
80
Error
no.
Display
Tests and/or remedial measures
Alarm contact
Error current
Autom.
cleaning start
PROFIBUS
status
Facty
User
Facty
User
Facty
User
PV
1)
Temp
Summary of Contents for Liquisys M CCM223
Page 109: ...Liquisys M CCM223 253 Technical data Endress Hauser 109...
Page 116: ...Appendix Liquisys M CCM223 253 116 Endress Hauser a0002639 en...
Page 120: ...Appendix Liquisys M CCM223 253 120 Endress Hauser...
Page 123: ...Liquisys M CCM223 253 Endress Hauser 123...
Page 124: ...www addresses endress com 71212626...