System integration
Liquiline Control CDC90
54
Hauser
7
System integration
7.1
Web server
7.1.1
Establishing the data connection
1. Start your PC.
2. First, set a manual IP address in the network connection settings of the operating
system.
3. Start the Internet browser.
4. If you use a proxy server to connect to the Internet:
Disable the proxy (browser settings under "Connections/LAN settings").
5. Enter the IP address of your device in the address line (192.168.0.1:8080/
CDC90.htm).
The system takes a few moments to establish the connection and then the web
server starts.
7.1.2
Operation
The menu structure of the web server corresponds to the onsite operation.
7.2
Fieldbuses
7.2.1
Connection
You can choose from the following communication options in the CDC90 control unit:
• Analog current output
Via AO modules in the CDC90 control unit. The settings can be performed via Web
server or local display.
• EtherNet/IP
• PROFIBUS DP (Slave)
• Modbus TCP (Client)
• Profinet
• OPC UA
Connection via Ethernet switch and fieldbus card. → 42
Ethernet cable assignment
RJ45
Stand. cable
Ind. cable
M12
1
Amber
TxD-
Amber
3
2
Amber/White
TxD+
Yellow
1
3
Green
RxD-
Blue
4
4
Green/White
RxD+
White
2
Assignment of M12 connection
M12
M12
1
Yellow
1
2
White
2