Electrical connection
Liquiline Control CDC90
52
Hauser
Terminal X1, bottom
Cable wire
L
L1, BN
PE
PE, GN-YE
N
N, BU
PE
PE
PE PE
PE
PE
1
3
N
4
3
L
PE PE
2
4
P
PE PE PE
PE
PE
PE PE
PE
PE
PE
A0035338
48
Terminal diagram of main supply voltage of actuator terminal X1 in the pneumatic control unit
6.9
Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions
and which are necessary for the required, designated use, may be carried out on the device
delivered.
‣
Exercise care when carrying out the work.
Individual types of protection permitted for this product (integrity (IP), electrical safety,
EMC interference immunity, Ex protection) can no longer be guaranteed if, for example:
• Covers are left off
• Different power units to the ones supplied are used
• Cable glands are not sufficiently tightened (must be tightened with 2 Nm for the
confirmed level of IP protection)
• Unsuitable cable diameters are used for the cable glands
• Modules are not fully secured
• The display is not fully secured (risk of moisture entering due to inadequate sealing)
• Loose or insufficiently tightened cables/cable ends
• Conductive cable strands are left in the device