Diagnostics and troubleshooting
Liquiline Control CDC90
92
Hauser
Item
Error message
Troubleshooting
304
Invalid measured value
1.
Check sensor power supply.
2.
Restart device under
Setup/Diagnosis/
Reset
.
304
Measured value not available.
‣
Check connection between sensor and
transmitter.
305
Delta slope warning, measurement
can still take place. The calibration
indicates a large slope change in
the sensor.
1.
Check sensor, replace if necessary.
2.
Check buffer or electrolyte, replace if
necessary.
3.
Repeat the calibration.
306
Sensor glass limit value exceeded.
Alarm impedance of glass
membrane too high.
1.
Check pH sensor and clean if necessary.
2.
Check setting for glass limit value and, if
necessary, correct under
Setup/Inputs/
Channel/Extended Setup/ Diagnostic
settings/Glass imp. (SCS)
.
3.
Replace sensor
306
Warning: Sensor glass limit value
exceeded. Impedance of glass
membrane too low. Measurements
can be performed until the alarm
occurs.
1.
Check pH sensor and clean if necessary.
2.
Check setting for glass limit value and, if
necessary, correct under
Setup/Inputs/
Channel/Extended Setup/ Diagnostic
settings/Glass imp. (SCS)
.
3.
Replace the sensor.
307
A major alarm has occurred in the
transmitter. Error message
available in transmitter.
307
A major warning has occurred in
the transmitter. Error message
available in transmitter.
308
A major warning has occurred in
the transmitter. Error message
available in transmitter.
309
Transmitter current error
310
Invalid measured value
1.
Check sensor power supply.
2.
Restart device under Setup/Diagnosis/
Reset.
400
Communication between control
unit and pilot valve manifold
canceled.
‣
Check connection between the devices.
401
An error has occurred during
calibration.
‣
Ensure that measurement remains stable,
calibration results are in the defined
warning limit, sensor is connected.
401
Calibration error has occurred
402
The connected sensor is not
supported for the calibration
function.
402
The sensor is not supported for an
automatic calibration.
403
No sensor selected for measuring
point.
403
No sensor selected for channel.
‣
Select a sensor for the channel.
404
The field device's calibration menu
is open.
‣
Close field device's calibration menu.
405
An error has occurred during
calibration.