Operation
Liquiline Control CDC90
88
Hauser
Function = Diagnostics
Function
Options
Info
Operating mode
Selection
• as assigned
• Namur M
• Namur S
• Namur C
• Namur F
Factory setting
as assigned
as assigned
If this option is selected, the diagnostic messages
which you have individually assigned to the relay
are output via the relay.
Namur M
...
Namur F
If you decided to use one of the Namur classes,
all the messages that are assigned to the
individual class are output via the relay. You can
also change the Namur class assignment for
every diagnostic message.
(
Menu
/
Setup
/
General settings
/
Extended
setup
/
Diagnostics settings
/
Diag. behavior
or
Menu
/
Setup
/
Inputs
/<Sensor>/
Extended
setup
/
Diagnostics settings
/
Diag. behavior
)
Attributed diagnostic messages
Operating mode
=
as
assigned
Read only
All the messages assigned to the relay output
appear on the display. You do not have the
option of editing the information here.
Outputting the status of a cleaning function
Function = Cleaning
Function
Options
Info
Assignments
Options
• None
• Depends on the type of
cleaning
Factory setting
None
Here you can specify how a cleaning function
should be displayed for the relay.
Depending on the cleaning program selected
(
Menu
/
Setup
/
Additional functions
/
Cleaning
)
you can choose from the following:
•
Cleaning type
=
Standard clean
Cleaning 1 - Water
,
Cleaning 2 - Water
,
Cleaning 3 - Water
,
Cleaning 4 - Water
•
Cleaning type
=
Chemoclean
Cleaning 1 - Water
,
Cleaning 1 - Cleaner
,
Cleaning 2 - Water
,
Cleaning 2 - Cleaner
,
Cleaning 3 - Water
,
Cleaning 3 - Cleaner
,
Cleaning 4 - Water
,
Cleaning 4 - Cleaner
•
Cleaning type
=
Chemoclean Plus
4x
Cleaning 1 - %0V
, 4x
Cleaning 2 - %0V
1)
%0V is variable text which you can assign in
Menu
/
Setup
/
Additional functions
/
Cleaning
/
Chemoclean
Plus
/
Output label 1 ... 4
.
Binary outputs
All the inputs are assigned for float switches and pressure switches
Hardware options, such as the "DIO" module with 2 digital inputs and 2 digital outputs or
fieldbus module "485" enable the following:
Via a digital output signal
– diagnostic states, point level switches, or similar states transmitted statically (similar to
a relay)
– the dynamic transmission (comparable to a non-wearing "analog output") of PFM
signals, e.g. to control dosing pumps