EM TEST
UCS 500 Series
Operating Manual
V 5.40
52 / 99
10.1.3.
User Test Routines
The user can program, save and recall his own specific test routines. The next pages shows the selection of the
functions.
User Test Routines
F1 : Change polarity after n pulses
F2 : Voltage change after n pulses by
V
F3 : Change the phase angle after n pulses by
a
F4 : IEC coupling after n pulses
F5 : ANSI A coupling after n pulses
USC 500N7 Firmware 9.00 and iec.control V 5.2.3.0 or later
F6 : ANSI B coupling after n pulses
USC 500N7 Firmware 9.00 and iec.control V 5.2.3.0 or later
F1
F2
F3
F4
F5
F6
F7
After selection the last used test parameters will be indicated on the display.
Customized test routines
The software controls user test routines according to the specification of the user. All limitations are the same as
defined under Quick Start.
F1
Change polarity after n pulses
After the release of the preselected number of pulses the
polarity is changed. The procedure always starts with
positive polarity and changes than to negative. The same
parameters as under Quick Start can be selected.
F2
Change test level V after n pulses by
V
The test voltage V is changed from V1 to V2. After the preselected
number pulses the test level is changed by
V until V2 is reached.
The same parameters as under Quick Start are selectable.
For the limitation of the max. admissible repetition rate the
higher value of V1 and V2 is valid.
F3
Change the phase angle A after n pulses by
A
The phase angle related to which the surge pulse is
released is changed from A1 to A2. After the preselected
number of n pulses the actual phase angle is changed by
A
until A2 is reached. The same parameters as under Quick
Start can be selected.