EM TEST
UCS 500 Series
Operating Manual
V 5.40
20 / 99
F7 Magnetic field correction factors
F1: Coil factor
Af
A/m
Range [0.200...9.99
step 0.01 ]
F2: Transformer factor
Tf
A/V
Range [0.020...9.999
step 0.001 ]
F3: Impedance factor
If
A/V
Range [0.200...1.00
step 0.01 ]
These values are delivered together with the necessary options to conducted magnetic field testing.
-
Coil factor depends on the type of antenna which is used for the test.
-
Transformer factor depends on the type of transformer which is used for the test.
-
Impedance factor depends on the type of surge generator which is used for the test. A generator with a
source impedance of 2
needs an impedance factor of approx. If = 0.5