4.6.
Setup procedure for Set voltage
The UCS 500 is able to control an external voltage source with a 0-10V analogue dc signal.
The 10V level corresponds to the max. output voltage of the connected voltage source. For the correct setting it
is necessary to know the max. output voltage of the source. For setting for ac or dc source the same procedure is
necessary.
Setup for adjust the external voltage variac to the UCS 500Nx
The voltage setting is a regulating procedure where one
winding of the motor-variac is approx. 1.9V. Additionally
the control needs some mechanical tolerance for stop
the step motor. Therefore the selected voltage is
normally in a tolerance of
3V
The motor-variac control has an internal deadband
where the voltage will not be adjusted.
The following procedure will guide the user through the Set Voltage procedure.
The setting for Umax and Un are listed in the boxes below. After each setting the user has to
press the ESC
button
for make the setting.
Step
Umax.
Un
CH
Remarks
A
--
--
PF2
Set the UCS 500 to PF2 mode for measure the variable voltage
B
100
100
PF2
The variac moves to his max. voltage.
( 10V dc at the 0-10V output)
Read Umax. voltage from the voltmeter.
( MV2616 approx. 257V)
C
Umax
230
PF2
Enter Umax. and set Un to 230V ( 115V)
D
Umax
100
PF2
Check the 100V at the voltmeter.
Tolerance
3V
E
Umax
0
PF2
Check the 0V at the voltmeter.
Tolerance < 2V
F
Umax
100
PF2
Check the 100V at the voltmeter.
Tolerance
3V
G
Umax
230
PF2
Check the 230V at the voltmeter.
Tolerance
3V
If the output voltage is not inside the tolerance or you like to make a better adjustment:
- Change the Umax voltage and restart the check procedure from step C to G.