EM TEST
UCS 500 Series
Operating Manual
V 5.40
51 / 99
Standard Test Routines
Page 3
(show parameters)
Surge
Generic EN 61000-4-5
5s
AUTO
10 Pulse
+ 1000V
L-PE
90grd
Counter
U=970V
000002
STOP STEP
I = 010A
000034
F1
F2
F3
F4
F5
F6
F7
- The counter shows the number of triggered pulses per actual test level as well as the number of all triggered
test pulses within the running test sequence.
- Pushing the function key F2 STEP will bring you into the next iteration sequence.
Manual test routine
Surge
Standard
IEC 61000-4-5
<- -> O O
Level3
2000V 10s 0 grd L-N
START
/- O V/A Rep L-N L-PE N-PE
F1
F2
F3
F4
F5
F6
F7
Within this test routine all standard parameters can be changed online during testing. This procedure therefore is
very easy and fast to use.
Example:
- Operating the cursor will increase/decrease the test level to the next level required in the standards
- By turning the inc knob the test voltage V resp. the phase angle A will be adjusted continuously. The blinking
circle shows which parameter can be changed. Pressing the function „O V/A“ will change between both
parameters.
- Pressing the function keys will activate immediately the related function.
All functions can be changed during the running test.