EM TEST
UCS 500 Series
Operating Manual
V 5.40
67 / 99
11.2.3.
User Test Routines
Page 2 (Selection)
Power Fail User test routines
F1 : Voltage variation as per IEC 61000-4-11
F2 : Change angle after n events by dA
F3 : Change duration after n events by dA
F4 : Inverse mode
F1
F2
F3
F4
F5
F6
F7
F1 Voltage variation as per IEC 61000-4-11
An external power source or motor driven variac is controlled
by a 0-10V control signal. The operator can select the time
per voltage level, the ramp up and ramp down of the voltage
change and the voltage levels itself.
Page 2 (show parameters)
Power Fail
Voltage Variation
Press
START
and the test routines begin to work.
Press
CHANGE
and the actual parameter can be changed.
V1
= 220 V
T1 =
10 s
V2
= 120 V
T2 =
1 s
V3
= 220 V
T3 =
10 s
Tf
= 2.0 s
Tr =
2.0 s
tri
= AUTO
n
=
00010
START CHANGE
F1
F2
F3
F4
F5
F6
F7
F2 Change angle after n events by
A
After n events the phase angle related to which the events
are released will change from A1 to A2 by steps of
A until
A2 is reached. The same parameters as under Quick Start
can be selected.
F3 Change duration after n by
td
After n events the duration of a single event will change from td1
to td2 by steps of
td until td2 is reached.
The same parameters as under Quick Start can be selected.
F4 Inverse mode
The inverse mode can simulate a phase control circuit, switching
power (voltage) on/off at a certain phase angle. The phase angle
is selectable in the range of 0-180°. The voltage will be switched
on/off in each half-wave. Inverse is only working in
U
mode.